microbiome research workstation expert
PUNMICRO Scientific
PRODUCT

Multi-Element Flame Photometer

Multi-element flame photometer
OVERVIEW

Five elements, one measurement pass

The multi-element model adds barium to the standard Na, K, Li and Ca set, so five elements can be measured and displayed at once. It is used where sodium and potassium have to be read alongside calcium and barium, for example in food nutrition, clinical samples, cement analysis and sugar processing.

  • Five elements measured and displayed simultaneously
  • Barium range from 5.0 to 3000 ppm, detection limit 1.6 ppm
  • Single-point, multi-point and simultaneous multi-ion calibration

Product features

Elements

Up to five elements measured and displayed at once; custom channels available.

Applications

Food nutrition (GB 5009.91-2017), clinical chemistry, cement quality control (GB/T 176-2017), sugar processing, pharmaceutical testing to the 2020 Chinese Pharmacopoeia, and glass analysis (GB/T 1347-2008).

Repeatability

Coefficient of variation below 1% across 20 consecutive samples at 100 ppm or lower.

Drift

Within 1% after a 30-minute warm-up.

Technical specifications

ElementsNa, K, Li, Ca, Ba (up to 5 simultaneous)
RangesNa 0.1–1000 · K 0.05–1000 · Li 0.1–1000 · Ca 1.0–1000 · Ba 5.0–3000 ppm
Detection limitsNa 0.02 · K 0.02 · Li 0.02 · Ca 0.3 · Ba 1.6 ppm
Resolution0.001
Repeatability< 1% CV (20 samples, ≤ 100 ppm)
Drift≤ 1% after 30 min
InterferenceNa/K/Li ≤ 0.5% (< 100 ppm)